PPC2-AF™ 1.01e PPC2-AF Do Not use with PPC2+!

 


Updates Include:
Ver 1.01 Rev e (2006 Nov 17)
CHG: Support new version of display (CU20025SCP-T200A) to eliminate flicker if display is replaced.
FIX: Gas selection menu under the fluid head menu would not default "He" when gas was previously set to "He"
FIX: Negative head entries would revert to positive values if a user simply pressed "ENTER" while viewing the negative value without changing it.

 

Ver 1.01 Rev d (2001 Sep 21)
CHG: Increase calibration adder (PA) and PAZTare edit limits from ± 999 to ± 9999.
CHG: Calibration multiplier limits from 0.9 - 1.1 to 0.8 - 1.2.

 

Ver 1.01 Rev c (2001 Apr 11)
FIX: Display conflict that could cause incorrect values of PA(z)net to be displayed following certain key sequences.
FIX: On-board calibration routine that in certain sequences of operation, when partial calibrations are run, may cause the PA(z)tare values to incorrectly be written as the PA values.
CHG: The 180 second "dwell time" in various functions has been reduced to 120 seconds.
CHG: The low point in the on-board calibration routines for ranges H1, H2 and H3 has been changed from 0.5 psia to 15 psia.


 


 

 

Download File:

 

Download Example Setup Database

FREE Download
ppc2af101e.bin (223 kb)

 

 

Instructions:

 

This software revision affects only the PPC2 AF on-board routines for calibration of the PPC2 AF's RPTs. It has no effect on regular operation. This new release is recommended for all PPC2 AF units, in particular those on which difficulty with calibration of the Lo Reference Pressure Transducer has been encountered. The DHI Service Department is systematically upgrading units returned to it for service to Ver. 1.01c.

Download the file.

 

Run Winflash and select the file downloaded as the file to write to the product.

 

Make sure the correct COM port of the product is connected to a RS232 port on the host PC.

 

Use a null modem cable to COM2 port of PG7000 Piston Gauge.

 

Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox.

 

If your product has
an model number ending in “-AF”,
it requires specific
“-AF” embedded software (see [USAF])

 

The updating of embedded software should only be performed after given specific instructions to do so from a DHI Authorized Service Center.

 

If the update isn't performed properly, your product will be inoperable until it is.
   