Embedded     [pressure ]    [flow]     [ USAF]

 

 Pressure

 

TITLE VER USE WITH  
RPM4™ 1.01h RPM4
donotuse
Note: If you have a special RPM4
model use these links:

RPM4-AD™
RPM4-E-DWT A70M/A7M-AF™

Contact DHI if upgrading from 1.01f


Updates Include:
Ver 1.01h (2009 Feb 09)
FIX: When powered on in HL (parallel) mode, the advanced Q-RPT characterization is not applied
for both IH and IL QRPTs.

 

Ver 1.01g (2008 Sept 08)
FIX: Full Q-RPT characterization is not being applied to pressure values.

 

Ver 1.01f (2008 Jul 30)
FIX: Unit change using the remote "UNIT=" command should not change the current measurement mode.
FIX: Cycling power with an unsaved AutoRange active results in odd behavior.
FIX: Absolute mode AutoZero via COM2 using an RPM4 results in an Pstd value of “0.000" being displayed instead of the reference RPM4's pressure.
FIX: RPM4 can boot to FATAL EXCEPTION BUS ERROR.
FIX: After the power up version screen, the screen goes blank for about 5 seconds before the main pressure measurement "Run screen” appears.

 

Ver 1.01e (2005 Nov 01)
FIX: Torr and mTorr unit conversions.

 

Ver 1.01d (2005 Apr 04)
Not for public release.

 

Ver 1.01c (2005 Mar 14)
ADD: pressure units "Torr" and "mTor"
CHG: Update the Head correction calculation to  ensures that the gauge pressure reads "0.000" after AutoZ with an Air Head applied.
FIX: Front panel menu locks when the UP or DN arrow key is pressed in a menu with selections that only span the bottom line.

 

Ver 1.01b (2004 Aug 25)
ADD: New G15K Q-RPT range. ADD: Rate is included in the reply to the "CONT" remote query.
FIX: Allow correct selection of pressure unit kcm2 FIX: The query character "?" is removed from the reply to query forms of the remote commands "*RST?", "ABORT?", "REMOTE?" and "LOCAL?".
FIX: RPM4 to stay in remote mode during and after use by a PPC3 until the power is cycled.
FIX: "CALAMB" command was missing the cal date in the reply.
FIX: Internal real time clock could not properly set the date past year 2009.
FIX: "GPIB" command now accepts correct address values from 1 to 30. Also, It now replies the correct address when changing the address.
FIX: Third field of the reply from the remote command "RPT" is now the RPT serial number instead of the RPM4's serial number.

 

Ver 1.01 (2003 May 12)
ADD: Support for use by PPC3 as an external pressure reference
ADD: Parallel measurement mode, "HL" Q-RPT functionality.
ADD: Differential mode
ADD: L2 and L3 commands
FIX: "COM1=" command could cause data corruption and a reset "ALL" on the next power cycle.
FIX: "COM1=" command reply may not be sent complete when setting a new baud rate
FIX: "LL=" command reply
FIX: Ensure that Auto-ranged UL values do not exceed the RPT's default range UL values

 


 

 

Download File:

 

Download Example Setup Database

FREE Download
101h.bin (275 kb)

 

 

Instructions:

 

Download the file.

 

Run Winflash and select the file downloaded as the file to write to the product.

 

Make sure the correct COM port of the product is connected to a RS232 port on the host PC.

 

Use a null modem cable to COM2 port of PG7000 Piston Gauge.

 

Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox.

 

If your product has
an model number ending in “-AF”,
it requires specific
“-AF” embedded software (see [USAF])

 

The updating of embedded software should only be performed after given specific instructions to do so from a DHI Authorized Service Center.

 

If the update isn't performed properly, your product will be inoperable until it is.
   