Ver 1.01e (2005 Nov 01)
FIX: Torr and mTorr unit conversions.
Ver 1.01d (2005 Apr 04)
Not for public release.
Ver 1.01c (2005 Mar 14)
ADD: pressure units "Torr" and "mTor"
CHG: Update the Head correction calculation to ensures that the gauge pressure reads "0.000" after AutoZ with an Air Head applied.
FIX: Front panel menu locks when the UP or DN arrow key is pressed in a menu with selections that only span the bottom line.
Ver 1.01b (2004 Aug 25)
ADD: New G15K Q-RPT range. ADD: Rate is included in the reply to the "CONT" remote query.
FIX: Allow correct selection of pressure unit kcm2 FIX: The query character "?" is removed from the reply to query forms of the remote commands "*RST?", "ABORT?", "REMOTE?" and "LOCAL?".
FIX: RPM4 to stay in remote mode during and after use by a PPC3 until the power is cycled.
FIX: "CALAMB" command was missing the cal date in the reply.
FIX: Internal real time clock could not properly set the date past year 2009.
FIX: "GPIB" command now accepts correct address values from 1 to 30. Also, It now replies the correct address when changing the address.
FIX: Third field of the reply from the remote command "RPT" is now the RPT serial number instead of the RPM4's serial number.
Ver 1.01 (2003 May 12)
ADD: Support for use by PPC3 as an external pressure reference
ADD: Parallel measurement mode, "HL" Q-RPT functionality.
ADD: Differential mode
ADD: L2 and L3 commands
FIX: "COM1=" command could cause data corruption and a reset "ALL" on the next power cycle.
FIX: "COM1=" command reply may not be sent complete when setting a new baud rate
FIX: "LL=" command reply
FIX: Ensure that Auto-ranged UL values do not exceed the RPT's default range UL values |