CalTool for Analog is only used with molbox1s that have an MFC option installed and with an MFC CB.
If you have one of these products, locate the calibration cable that was shipped with the product and
obtain a DMM for calibration. Install and run CalTool for Analog, then follow the instructions in CalTools
help menu.
In general, follow these steps:
1. Select the [Setup] menu choice.
2. Specify the remote interface to the product (molbox1 or MFC CB) and for the DMM.
3. Enter the desired voltage and current test points. The default values are recommended.
Press [Accept] when ready.
4. Press the [Begin Test] button at the bottom of the main display.
5. Several prompts will follow to determine the type of calibration to perform (voltage, current or both).
6. Connect the product to the DMM as requested by CalTool and follow the program prompts.
7. CalTool will acquire data from the product and the DMM and determine new calibration coefficients.
Options are provided to write this information to the product and to generate a basic calibration report. |