GFS Tools is a PC based application used to operate the GFS 2102 Gravimetric Flow Standard. |
| summary |
GFS Tools is the PC based interface to the GFS 2102. All necessary features are included to log data for up to 2 DUTs in comparison to the GFS from 1 sccm of N2 to 10 slm of N2. Options are provided to set manual or automatic flows using an MFC and to define a test sequence of flow points to fully automate a calibration. GFS Tools also provides maintenance functions required to ensure consistent functionality of the GFS 2102.
GFS Tools is specifically designed to operate a GFS 2102. This program cannot be used without complete GFS 2102 hardware. |
updates include |
Rev 1.00f (2007.11.29)
ADD: Improved error handling on instantaneous mass rolling best fit.
FIX: The maximum cylinder probe temperature offset value is not used.
|
Rev 1.00e (2007.03.08)
ADD: Support the option to tare a molbox during a GFS AutoZero
ADD: A new configuration option is available to determine how many balance readings are used to determine the mass for any given GFS sample.
ADD: Support for Win XP restricted users.
CHG: A value of 0 will defeat the rolling avg configuration settings
CHG: Delete files to the Recycle bin instead of a permanently deleting them.
|
Rev 1.00d (2006.11.16)
ADD: Custom data file naming is available in [Tools], [Options], [Data File].
ADD: Support extended 64K molbloc EEPROM calibration information.
CHG: Removed options to regulate flow to the GFS outputs.
CHG: GFS and DUT *.sdf files are sequenced according to the active test data file name.
CHG: Tests that regulate to DUT2 will regulate to DUT1 when only 1 DUT is calibrated.
CHG: Changed the main form and shortcut icon.
CHG: Changed default balance sample period to 100ms instead of 300ms.
FIX: The DUT Reynolds number is stored using "RE" as the unit of measure.
FIX: The gas compressibility factor is used in an iteration to determine the pressure inside the GFS reference gas cylinder.
FIX: Automated tests will now get a flow Ready indication when regulating to the second DUT.
FIX: The data file only contains data from accumulations actually run.
FIX: The air-buoyancy correction is removed prior to determining the actual change in mass of the cylinder.
FIX: The test averaging time is based on the averaging time value in the Test Definition.
FIX: Hold and stability unit selection fields in the Test Definition Editor, are in alignment with their numeric entry fields.
FIX: The template selected in the active test definition is used for report generation when the test is complete.
FIX: Faulty cursor highlighting.
FIX: Provide option to manually close the flow isolation valve.
|
|